Agilent DNA 7500 and DNA 12000 15
Agilent DNA 7500 and DNA 12000 Assay Protocol 5
Loading the Ladder and the Samples
Loading the Ladder and the Samples
1 Pipette 1 µl of the yellow- capped DNA ladder vial
(yellow ) in the well marked with the ladder
symbol .
2 In each of the 12 sample wells pipette 1 µl of
sample (used wells) or 1 µl of deionized water
(unused wells).
3 Set the timer to 60 seconds.
4
Place the chip horizontally in the adapter of the IKA vortex mixer and
make sure not to damage the buldge that fixes the chip during vortexing.
5 Vortex for 60 seconds at 2400 rpm.
6 Refer to the next topic on how to insert the chip in the Agilent 2100
bioanalyzer. Make sure that the run is started within 5 minutes.
1 µl ladder
1 µl sample
CAUTION
Wrong vortexing speed
If the vortexing speed is too high, liquid spill that disturbs the analysis may occur for
samples generated with detergent containing PCR buffers.
Reduce vortexing speed to 2000 rpm!
NOTE
For optimal results, samples should be of pH 6 to 9 and should not have an ionic content
greater than twice that of a typical PCR buffer.